RNA can be quantified using absorption of light at 260 and 280 nm (A260/280). Ideally, this ratio should be close to 2 for high quality nucleic acid. However, for retinal RNA, we typically get values more like 1.8 and values as low as 1.6 still give useful real time RT PCR results (consistent with other samples).
Warm up the spectrometer with the UV light on for at least 15 min.
Make a 1:10 dilution for each RNA saple. If you are using a 50 ul cell, 6 -> 60 ul water would be about right.
Read a blank by putting 60 ul of water into the cell. Then read each of your samples, rinsing the cuvette with water in between samples. I don't reuse this RNA as I don't want any contamination.
A crude way to calculate the concentration from the absorption at 260 nm is to use Beer's law : A = e C l
where A is the measured absorption at 260 nm
e is the RNA extinction coefficient (25 ul / ug / cm)
C the RNA concentration
and l is the pathlength (1 cm)
From this C(meas) = A / (e l) = A * 40 ug/ml. However, this is the concentration of the diluted solution in the quartz cuvette. The actualy RNA concentration C(sample)=C(meas) * dilution factor. In the above case the dilution factor is 10.
For example, if we measured A=0.08, then C(meas)=0.08 * 40 ug/ml = 3.2 ug/mlA more precise way to get RNA concentration (and take account of protein absorption which is important when A260/A280 is much less than 2) is to use the method described in Glasel, Biotechniques 18: 62-63 (1995). I have an excel spreadsheet which does this calculation if you want to try it.